Features:
- New technology of non-invasive arterial blood pressure measurement Unicuff ™:
- High precision measurements in all clinical situations (arrhythmia, low pulse wave)
- Maximal comfort for the patient and minimal influence on hemodynamics
- Swift measurement (20-30 seconds)
- Measurement stability and update on cuff deflation
- SpO2 and ECG Modules with improved calculation algorithms, high stability and resistance to artifacts
- Two temperature measurement channels for ease of patient condition monitoring
- 10,4” color display with high resolution. High image contrast, large viewing angle
- Adjustable display with large font “Top-DIGIT™” mode
- Easily available and inexpensive accessories
Specification:
ECG Channel:
- Monitoring of I, II, III, aVL, aVR, aVF
- Manual selection of cardio signal strengthening coefficient
- Measurement range: 12-300 QRS complexes per minute
- Digital display of ST-segment displacement relative to isoline
- Display speed 12.5, 25, 50 mm per second
- Protection from signal interference from electrosurgical instruments
- The device can be used together with defibrillator
- Basic arrhythmia analysis
Pulse oximetry channel:
- Digital display of pulse rate and SpO2
- P hotoplethysmogr amm
- Automatic photoplethysmogramm scaling
- Measurement range of SpO2 from 60% to 100%
- Pulse rate measurement range: from 5 to 300 beats per minute
- Error limits when measuring SpO2 in the range of:
90-100 % ±2
60-89 % ±3
- Error limits when measuring pulse rate in the range of:
25-99 ±1 beats/minute
100-220 ±2 beats/minute
Tonometry channel (non invasive arterial blood pressure measurement):
- Digital display of arterial pressure: systolic, diastolic, average
- Pressure measurement range : 0-300 mmHg
- Absolute measurement error ±3 mmHg
- Time to measure: no more than 40 seconds
- Adjustable time interval between measurements: from 1 to 60 minutes, with a 1 minute control step
- Adult and child measurement modes
- Method of measurement: oscilloscopic
Respiration channel:
- Digital display of breathing frequency;
- Respirogramm – breathing graph display in the breathing frequency range between 6 and 60 breathes per minute
- Breathing rate measurement range: 0-160 breathes per minute
- Absolute measurement error in the whole range: ± 3 breathes per minute
Thermometry channel (2 channels – skin and cavity temperature):
- Digital temperature display
- Measurement range: 20-43 degree Celsius (68 – 109.4 degree Fahrenheit)
- Possible measurement error ±0,1 degree Celsius;
Alarms:
- Three-tier alarm system with audible and visual (in various colors) warnings alerts to changes in measured parameters
- Adjustable alarms: heart rate, SpO2, temperature, systolic pressure, diastolic pressure, breathing rate, ST-segment displacement
- Non-adjustable alarms: asystolia, blood pressure measurement error, sensors of SpO 2 or ECG or temperature not installed properly, SpO2 sensor malfunction, SpO2 sensor not plugged in, ECG low voltage, photoplethysmogramm low signal, NIBP low signal, respiratory standstill, lead removed
Data storage:
- Trends are shown as separate trend graphs with different color for each monitored parameter and for alarms
- All graphs are displayed in real time in a single window
- Operator can select which graphs to display
- Data is saved with 1 minute steps, and past data can be shown anytime, with the trend displayed as a table
- Basic memory volume allows saving 48 hours of measurement in 1 minute steps
- Data storage is immune to power loss
- Graphic trends : 1,5; 3; 6; 12; 24; 48 hours
- Data saving is performed automatically
Recordable trends:
- Heart rate
- SpO2
- Arterial blood pressure
- Temperature (2 channels)
- Peripheral circulation
- Breathing rate
- ST-segment displacement size
User interface:
- Color TFT screen with active backlight (the device can be used in dimmed rooms or in a car or plane)
- Display size 10.4 inches ( 26 centimeters)
- Up to five waveforms can be displayed simultaneously
- Screen resolution 640x480 pixels
- Can be plugged into a local network
- Heart rate sound indication
- Integrated three - tier alarm system
- Convenient “shuttle” manipulator allows fast and convenient operation of the device
Accessories (standard set):
- Three-lead ECG cable 3 meter – 1
- Set of disposable leads neonatal – 60 sets
- Set of disposable leads adult – 60 sets
- Pulse oximetry clip-on sensor adult with 3 meter ( 9.84 feet) cable – 1
- Pulse oximetry clip-on sensor neonatal with 3 meter ( 9.84 feet) cable – 1
- Skin temperature sensor with 3 meter ( 9.84 feet) cable – 1
- Cavity temperature sensor with 3 meter ( 9.84 feet) cable – 1
- Pneumatic supply line for non-invasive arterial blood pressure measurements 2 meter ( 6.56 feet) long – 1
- Arterial blood pressure measurement cuff adult – 1;
- Arterial blood pressure measurement cuff adolescent – 1
- Arterial blood pressure measurement cuff neonatal – 5
Power supply :
- The device can be plugged into 220V – 50GHz alternating current power supply
- In-built internal battery
- Battery is recharged automatically
- The device can run on battery for no less than 2 hours
- Power consumption: no more than 35 VA
Size and weight:
- Device dimensions 300 х 180 х 230 ± 2 mm (11,811x7,086x9,055 inches ± 0,078 inches)
- Weight: no more than 4,9 kg ( 10,80 pounds) with battery and power supply unit
Monitor MPR 6/4-03 TRITON is compatible with Centralized Monitoring System SCM TRITON.
|